北京菲腾光电科技有限公司

 
产品详情
Fibered Fourier-Transform Infrared Spectrometer
分享到:

Fibered Fourier-Transform Infrared Spectrometer

产品详情
FT-IR Rocket
Fibered Fourier-Transform Infrared Spectrometer

The FT-IR "Rocket" is a robust, small-footprint, lightweight and high performance FT-IR spectrometer. The FT-IR Rocket can be used in a fiber connected version or a free space version. The device is very useful for IR measurements that needs fiber coupling or any other special arrangements like for IR light source characterization.
FT-IR "Rocket" offers high resolution and sensitivity thanks to the use of thermoelectrically cooled MCT (Mercury-Cadmium-Telluride) detectors. Its permanentely aligned interferometer and its solid-state reference laser ensure that the highest performance is delivered at every scan. The system has high stability in both wavelength and intensity scales, making it an ideal tool for highly reproducible chemiometrical analysis. The FT-MIR and FT-IR are also used by many of our customers for characterizing infrared lasers, who are using is an an Optical Spectrum Analyzer (OSA).

Standard models of the FTIR-Rocket

FT-MIR Rocket 2-6μm

FT-IR Rocket 2-12μm

The FT Rocket infrared spectrometer is proposed in 2 standard versions: The FT-MIR 2 - 6μm, FT-MIR 1.5 - 8.5μm and the FT-IR 2 - 12μm. Both offer a resolution of 4cm-1. The main differences lie in the beamsplitter material (CaF2 or ZnSe) and the sensitivity of the detector: the narrower band 2-6μm detector has a much better sensitivity. On request we can fit the detector to your needs.

  • Compact (world Smallest FTIR)

  • High resolution 4cm-1

  • Peltier cooled MCT detector

  • 3 spectral ranges: 5000-1700cm-1 (2-6μm), 6700-1200cm-1 (1.5 - 8.5μm) or 5000-830 cm-1 (2.5-12μm)

  • USB connectivity

  • Cost effective

ModelsFT-MIR 2-6
FT-MIR 1.5-8.5
FT-IR 2-12
Beam-spliter materialCaF2CaF2ZnSe
Spectral Range [cm-1]5000 - 17006700 - 12005000 - 830
Spectral Range [μm]2-61.5-8.52-12
Detector Type
D* [cm Hz1/2W-1]

MCT (2-TE cooled)

>1x1011

MCT (4-TE cooled)

>8x109

MCT (4-TE cooled)

>2.5x109

Signal-to-Noise (ratio between dark signal and peak intensity produced by 20W ARClight MIR)> 1:5'000> 1:3'000> 1:3'000
Recommended fiber
(usable wavelength range)

CIR (chalcogenide) fibers
1-6μm

CIR or PIR depending of the applicationPIR (polycrystalline) fibers
3-18 μm

Common specifications

Interferometer type

Permanentely aligned, double retro-reflector design

Resolution [cm-1 ]
(unapodized)

4

Wavenumber repetability

<10ppm
Scan frequency1 spectrum/second
(in continuous aquisition mode)
Control laser

Temperature-stabilized
solid-state laser @850nm

A/D Converter24 bit
Operating temperature10°C-40°C
Free-space inputAperture Ø: 12.7mm
Acceptance full angle: 3.2°
Removable fiber-couplersConnector: SMA 905
Fiber core Ø: up to 1mm
NA=0.25 (lensed coupler) or NA=0.3 (reflective coupler)
Power requirement12V / 8W
Communication InterfaceUSB 2.0
Software Interface

Windows XP/7/8/10
API for controlling the instrument via our DLL

Dimensions180mm x 160mm x 80mm
Weight1800 g

Wavelength resolution